|
Power Contacts/Connectors, Part III: Current Rating 1
By
Dr. Robert S. Mroczkowski, Bishop
& Associates Inc.
A number of issues
need resolution in order to perform a current rating test program.
Both practical and philosophical issues should be taken into
consideration as you embark on this process. We first discuss the
practical issues.
The relevant document for this discussion is EIA 364 D: TP-70B,
temperature rise vs. current for electrical connectors and sockets
(June 1997). As noted earlier and referenced in the title of TP-70B,
current rating is based on the T-rise of a connector under current
flow. T-rise is defined as the difference between the ambient
temperature of the test environment and the temperature of the
hottest point—the hot spot—on the energized contact. The most common
T-rise criterion is a 30-degree Centigrade difference as specified
by Underwriters Laboratories, a standards organization primarily
dedicated to safety issues.
TP-70B discusses temperature measurement issues in some detail.
Measurement of the ambient temperature is straightforward; the
measurement of the contact hot-spot temperature is more involved.
Two cases are discussed. In the first case, the contact hot spot is
accessible for a thermocouple or infra-red temperature measurement.
The second occurs when the hot spot is not accessible. Details for
such a measurement are still “to be determined” in TP-70B, but some
comments about this measurement are just ahead.
Accessible contact temperature measurement is defined in some detail
in TP-70B. The key points are thermocouple size and placement, and
IR measurement capabilities. Contacts within a housing are not
necessarily inaccessible. TP-70B describes procedures to provide
thermocouple access to such contacts. IR access is not discussed.
Inaccessible contacts are more problematic. The main challenge is
developing a correlation between the temperature at the hot spot
with some point on the contact which is accessible. The correlation
can be derived from either thermocouple or IR measurements using
contacts that are not mounted in the connector housing. The expected
rated current can be applied to such contacts, mounted in air, and
the steady-state temperatures at the hot spot and accessible point
can be measured. The difference in those temperatures is then
assumed to be the same when the contacts are inserted in the
housing. This, of course, is not necessarily the case, but because
all T-rise measurements are comparative, ensuring that all
qualification testing is done under identical measurement protocols
will lead to consistent evaluation of relative product performance.
The resolution or agreement on temperature measurement protocols are
generally straightforward.
A second testing issue involves the definition of steady-state
T-rise. This issue is addressed in detail in TP-70B. Section 4.1.3.1
is quoted in full:
“Maintain current
until thermal stability is achieved on all specimens. Thermal
stability shall be achieved when the temperature rise of a minimum
of three consecutive readings taken at five-minute intervals maximum
does not differ by more than ± 1°C (1.8°F) for each thermocouple
being used for monitoring. Thermal stability applies to all current
applications for all test methods described herein.”
Procedures for
achieving stability by gradually increasing applications of current
are also included.
Two philosophical issues are also important. The first is how to
mount the contact to perform the T-rise measurement, in free air or
in the housing. The second is the state of the contact in terms of
any conditioning steps. The extremes of this decision are a contact
as manufactured, that is, with no conditioning, to be referenced as
Out Of the Box (OOB), and a contact conditioned to an End Of Life (EOL),
state.
Consider first, free air vs. mounted in the housing. The T-rise of a
contact will be determined by the balance of Joule (I2R)
heating, which increases the temperature of the contact, and thermal
dissipation, primarily through conduction and convection, which
decreases the temperature of the contact. Clearly, free air
measurements will provide a greater degree of heat dissipation
compared to an in-housing measurement. However, application
conditions use contacts in housings, so in-housing measurements are
typically preferred.
Now, let’s look at decisions relevant to contact conditioning.
Testing OOB contacts verifies that the design objectives of the
contact have been met. But such measurements provide no indication
of the performance stability of the contacts in the field. Thus,
some variant of an EOL conditioning is more appropriate prior to
exposure to a current rating test program. Two examples illustrate
the difference between OOB and EOL.
First, a product is generally designed to have a specific
application life in the field; this is the “L” in EOL. Say that L is
10 years, a high quality white goods product, for example. It is
well known that contact normal force is a critical parameter in the
performance and, in particular, to the stability of a connector
system.
It is also well known that contact normal force decreases with time
through stress relaxation, a time and temperature dependent process.
For the white goods application under consideration, the application
temperature will be approximately room temperature. A 10-year
lifetime at room temperature can be simulated by a shorter time at
an elevated temperature, thus a 10-year life may be simulated by a
few months of testing at a higher temperature. An OOB test will not
provide any information about this potential degradation mechanism.
Second, a connector is typically rated for some number of mating
cycles. This parameter is intended to validate the integrity of the
contact finish against wear processes during mating. The number of
mating cycles corresponding to EOL is application dependent. In many
applications, a few mating cycles will be sufficient. In others,
higher numbers may be appropriate. Consider, for example, the USB
printer socket on a laptop computer. Some users may use their laptop
primarily at home, so their printer socket will be exposed to only a
few or a few tens of mating cycles. On the other hand, a field
salesperson will be using the laptop on the road and at home, and
may accumulate several hundred mating cycles over a few years. What
is EOL for a printer USB socket? Is it appropriate to test for a
typical—whatever that means—user, say, several tens of cycles? Or,
is it more appropriate to test to the design EOL, say 500 mating
cycles. Mating cycles can be performed quickly and at low cost, so
the issue here is not the test conditioning, the issue is
performance.
Different choices of EOL can be rationalized. The important point,
particularly from a user perspective, is to make sure that the test
programs are consistent. A stress relaxation or heat age test of 10
days at 75 degrees Centigrade is much less demanding than a 30-day
test at 125 degrees Centigrade. Just to put in a new consideration,
more is not necessarily better. Depending on the contact system
dependence on the housing, a 30-day, 125-degree-Centigrade heat age
test may not be appropriate due to contributions attributable to
housing degradation. That, however, is a different article.
For this article, the final point to consider is what exposures that
simulate EOL, however defined, are appropriate. While the list of
exposures can be very long, I suggest that there are four
indispensable exposures: durability, heat age, corrosion, and
mechanical stability. We’ll talk about them next time.
 |
Dr. Robert S. Mroczkowski
Director Technology, Bishop and Associates Inc.
IIn 1998, Dr. Mroczkowski founded connNtext associates, a firm
providing consulting services in connector applications to the
electronics industry. Dr. Mroczkowski has more than 30 years
experience in the electronics industry. He joined AMP Inc. in
1971. While at AMP, his responsibilities included consulting on
connector design, materials, and reliability concerns, and he
provided an interface to AMP customers on these issues. In 1990
he joined the AMP Advanced Development Laboratories, where he
developed microstrip cable connectors and a new microcoaxial
connector for medical ultrasound diagnostic equipment. Dr.
Mroczkowski retired in 1998 as an AMP principal. He is the
author of the
McGraw Hill Electronics Connector Handbook, has contributed
chapters on connectors and interconnections to a number of
packaging handbooks, and written more than 20 technical papers.
He holds seven patents. In 1997, Dr. Mroczkowski received the
Lifetime Achievement Award of the International Institute of
Connector and Interconnection Technology.
He holds a bachelor’s, master’s, and doctorate of science
degrees in physical metallurgy from the Massachusetts Institute
of Technology.
Contact Dr. Mroczkowski at
ConnNtextassoc@aol.com.
|
|